Toluene and the Intoxilyzer 5000: no response to concentrations found on human breath. |
作者: |
GLOVER-PL (North Carolina Dept of Health & Human Services, Raleigh, NC, USA) |
关键词: |
ALCOHOL-; 7347-; BLOOD-ALCOHOL-CONTENT; 1553-; DETECTION-; 9115-; BREATH-TEST; 1542-; EVALUATION-ASSESSMENT; 9020- |
摘要: |
The response of the Intoxilyzer 5000 (series 66, 3 filter) to known amounts of ethanol vapor, toluene vapor and a mixture of ethanol and toluene vapors was determined. The ethanol concentration approximates a Breath Alcohol Concentration (BrAC) of 0.08 gm/210 L. The toluene concentration used was approximately 5 times that reported to be found on the breath of test subjects 30 minutes post exposure. There was no response by the Intoxilyzer to the toluene alone and there was no added response to the ethanol/toluene mixture response over that of ethanol alone. For the covering abstract see ITRD E106992. |
总页数: |
PROCEEDINGS OF T2000 - 15TH CONFERENCE ON ALCOHOL, DRUGS AND TRAFFIC SAFETY, HELD 22-26 MAY 2000, STOCKHOLM, SWEDEN. 2000. pp5 |